UDK 139 Semi-Automatic Kjeldahl Distillation Unit

You Are Here:

»Home »Products » UDK 139 Semi-Automatic Kjeldahl Distillation Unit

Semi-automatic Kjeldahl distillation unit for your analyses concerning different applications such as determining ammoniacal nitrogen, protein determination, nitrogen content (Kjeldahl or direct alkaline distillation), nitric nitrogen (after reduction/Devarda), phenols, volatile acids, cyanides and alcohol content.

 

Automated Operation

  • Programmable water, sodium hydroxide addition and distillation time
  • Selectable steam generation output level 10% – 100%
  • Auto removal of residues from sample tube

 

High Tech

  • 3.5 inch color touch screen
  • Library of 10 customizable methods
  • Choice of 6 languages
  • Mouse & printer USB connections
  • No loss of nitrogen: distillate temperature stays below threshold
  • Accurate dosing of reagents with high precision pumps

 

Outstanding Reliability

  • Maintenance-free, resistant, patented steam generator
  • Patented titanium condenser for unmatched performance
  • Durable technopolymer splash head
  • Chemically-resistant technopolymer housing

 

Operator Safety

  • Lever operation simplifies tube insertion / removal (all sizes)
  • Protection screen for distillation chamber
  • Drip tray collects splashes
  • Zero pressure steam generator

 

Ecology & Standards

  • Performs to AOAC, EPA, DIN and ISO standards
  • Incorporates TEMS™ technology*
    Automated water supply control reduces consumption

 

Technical Data

 

Structure Corrosion-resistant technopolymer
Display 3.5" color touch screen
Time related to start for Devarda alloy analysis 0 – 99 min
Protocol library 10 customizable methods
Reproducibility (RSD) ≤ 1 %
Recovery ≥ 99.5 % at nitrogen levels between 1 and 200 mg N
Detection Limit ≥ 0.1 mg N
Power 2100 W
Weight 26 kg / 57.3 lb
Dimensions (WxHxD) 385x780x416 mm (15.2x30.7x16.4 in)

PERFORMANCE

 

Distillation time: 4 minutes to collect 100 ml of distillate
Sodium hydroxide addition: automatic
Water addition automatic
Sodium hydroxide volume 0 - 150 ml
Water volume 0 - 200 ml
Distillation time: settable
Distillation residues removal: automatic
Steam flow regulation from 10 to 100 %
Tap water consumption from 0,5 l/min at 15 °C – from 1 l/min at 30 °C
Connections 2 x USB for printer, mouse