A versatile and cloud enabled elemental analyzer able to perform precise carbon and nitrogen analysis in 4 to 5 minutes. It is the ideal solution for high throughput offering complete safety to operators.
Versatile
- Operates with Argon and Helium as carrier gas without any hardware modification
- Preinstalled methods with element selection (C,N and C/N)
- Controllable from PC by the powerful CNsoft™ software
- Clear and detailed reporting software
- Data storable in multiple formats for LIMS or PC
Productive
- Flash determination of nitrogen content
- Determination of TC, TOC (after acidification), TIC, TN and Carbon/Nitrogen Ratio
- Automated analysis of preloaded samples (117 positions)
- Totally unsupervised operation
- Results produced 24/7/365
Precise
- NDIR Detector (Non Dispersive Infrared) to detect Carbon concentration
- LoGas™ TCD with no reference gas required, for Nitrogen content
- Unmatched LOD: Carbon 0.01 mgC, Nitrogen 0.001 mg N (using He) or 0.01 mg N (using Ar)
- RSD less than 0.5% with EDTA standard
- Conforms to Good Laboratory Practice standard
- Leak tests (zone or total) ensure optimum performance
Economical
- Exceptionally low cost per analysis
- Premium quality consumables
- Low energy consumption with Stand-by and Sleeping mode
- Lowest operator costs
Environmentally Friendly
- Minimum possible waste & residues
- Optimized life of consumables
- Incorporates TEMS™ technology*
Cloud-Enabled Instrument
- Effortlessly connect your instrument to the cloud in complete safety
- Real time monitoring and notification via VELP Ermes Cloud Platform
- Immediate Software Updates (OTA)
- Access to your data via PC, smartphone and tablet
- Create and share reports, track trend and get useful insights
Technical Data
Method of analysis | Cabon/Nitrogen determination according to combustion method |
Duration of analysis: | 4/5 minutes |
Detector | N= TCD (Thermal Conductivity Detector) C= NDIR Detector (Non Dispersive Infrared) |
Sample weight: | max 1g |
Autosampler capacity: | up to 4 discs, 30 positions each (up to 117 samples) |
Reproducibility (RSD): | < 0.5% for EDTA standards approx. 100 mg |
Recovery | > 99.5% |
Detection Range C: | ppm - 150 mgC |
Detection Range N: | ppm - 200 mgN |
Carrier gas: | Helium or Argon |
Nitrogen Detection limit (Absolute values): | 0.001 mg N, with Helium 0.01 mg N, with Argon |
Carbon Detection limit (Absolute values): | 0.01 mgC |
Combustion temperature: | 1030 °C / 1886°F |
Carrier gas (He/Ar): | purity 99.999% (grade 5.0) |
Oxygen (O2): | purity 99.999% (grade 5.0) |
Carrier gas (He/Ar) pressure: | 2 bar |
Oxygen (O2) pressure: | 2 bar |
Connectivity | Cloud via LAN or Wi-Fi |
Interfaces | USB; RS232 |
Power | 1400 W |
Power supply: | 230 V / 50 - 60 Hz |
Weight | 54 kg |
Dimensions (WxHxD) | 655x510x410 mm (H 690 mm with autosampler) |