SL 14K

You Are Here:

»Home »Products » SL 14K

The new calibration baths are perfectly suitable for applications in calibration laboratories and conform to the requirements specified by DIN ISO 9001. The instruments provide highest temperature stability and temperature uniformity. An optional precision reference sensor with integrated measuring electronics and RS232 connection is available.

 

Your advantages

  • VFD COMFORT DISPLAY
  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Early warning system for low liquid level (DBGM 203 06 059.8)
  • Adjustable high temperature cut-out, visible via display
  • RS232/RS485 interface for online communication
  • Integrated programmer for 6 x 60 program steps
  • Highest temperature stability to ±0.005 °C (±0.02 °C at +200 °C)
  • Temperature uniformity better than ±0.01 °C
  • Optional Pt100 sensor for reference temperature measurements
  • Temperature chamber with overflow at maximum liquid level

 

Technical Data

Order No. 9352514
Model series HighTech
Category Calibration Baths
Working temperature range (°C) 50 ... 300
Temperature stability (°C) ±0.005
Setting / display resolution 0.01 °C
Integrated programmer 6x60 steps
Temperature Display VFD, LCD
Heating capacity (kW) 3
Pump capacity flow rate (l/min) 22-26
Pump capacity flow pressure (bar) 0.4-0.7
Pump capacity flow suction (bar) 0.2-0.4
Bath opening / bath depth (W x L / D cm) Ø = 12 / 31 cm
Filling volume (liters) 14
External Pt100 sensor connection integrated
Digital interfaces RS232
Optional Profibus
Ambient temperature 5 ... 40°C
Dimensions W x L x H (cm) 22 x 46 x 61
Weight (kg) 20
Classification according to DIN12876-1 Classification III (FL) according to DIN 12876-1
Included with each unit Including bath cover with 7 openings . Temperature stability at +200°C: ±0.02°C
Bath cover Integrated

Characteristics

 

A perfect view
Ample, easy to read VFD Comfort display for simultaneous display of 3 values, warning functions, high temperature cut-off, pump stages (resolution 0.01 °C)
Additional plain text information
Comfortable LCD dialog display for interactive operation with plain text display
Pump stage and liquid level
Backlit indicator for selected pump stages and filling volume on Presto® PLUS, Magnum 91 & Forte HT
Comfortable and detailed
Comfortable keypad with additional menu functions for pump stages, calibration, control parameters, programmer, warnings, etc.
For perfect results
Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.005 °C internal, <±0.05 °C external
Full control
Temperature Control Features’ for individual optimization, access to all important control parameters, additional settings for band limit, limits, Co-Speedfactor etc.
Highest measuring accuracy
Absolute TemperatureCalibration’ for manual compensation of a temperature difference, 3-point calibration
Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages
Highest measuring accuracy
Absolute TemperatureCalibration’ for manual compensation of a temperature difference, 3-point calibration
Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Connection compliant to standard
RS232/RS485 dual-interface for serial data transmission according to EIA-485 industry standard (2-wire bus technology), upgradable with Profibus DP
Optimal program control
For the execution of time and temperature dependant profiles, 6 temperature profiles with 60 steps max., with real time clock
Automatic control of operating time
Electronic countdown-timer function for timer-programmed unit shut-down, standby mode after programmed time expires
Early warning system for low liquid level
Maximum safety for applications, optical and audible alarm, allows user to refill bath fluid before the unit shuts down
Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function
Enhanced protective functions 
Maximum safety, adjustable high temperature cut-off or dry-running protection, additional display of setpoints permits easy and precise adjustments
For flammable bath fluids
Classification III (FL) according to DIN 12876-1
Consistent cooling capacity
Easily removable venting grid for quick and easy cleaning
100 % Cooling capacity
Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Condensation and ice protection
A heated cover plate prevents condensation or ice build-up in the bath
   

 

Other characteristics

Optional: Electronic module with analog connections for external programming, temperature recorder, standby, alarm (Order no. 8 900 100).

Technical changes without prior notification. Images may deviate from the original.